CT Training – CTA Exam
CTA EXAMS
CTA stands for Computed Tomography Angiography – CTA is a CT scan that uses a high-speed contrast injection and proper timing to scan vessels/organs of a specific area, while the contrast density is at its peak.
Most common CTA exams are:
- Abdominal CTA (See Figure 4-1)
- PE Study (Pulmonary Embolism) (See Figure 4-2)
- Circle of Willis (C.O.W) (See Figure 4-3)
- Run off (See Figure 4-4)
- Aorta
- Carotid




The key to a successful CTA exam is timing the scan properly to get the bolus of contrast in the anatomy of interest at the time the scan is performed.
First let’s talk about the contrast injection.
To achieve a quality CTA exam the contrast must be injected at a high rate of speed, usually 4 – 5 cc per second. The purpose of the high-speed injection is to keep a bolus of dense contrast moving through the vessels/organs at the time the scan is acquired.
Many facilities will also inject a bolus of saline after the contrast has finished injecting. The purpose of the saline injection is to push out the left-over contrast from the injection line and is thought to keep the contrast bolus intact until the scan is complete.
Type of Contrast – CT Training – CTA Exam
Many facilities will use a contrast that has a higher concentration of iodine for CTA exams. For example, they may use Isovue 300 for standard CT exams, but will use Isovue 370 for CTA exams.
The number on the contrast bottle represents the concentration of iodine within the solution, Isovue 370 will contain more iodine then Isovue 300. The use of higher concentrations of iodine for CTA’s helps to better opacify the vessels of interest.
IV Selection
Since the contrast is being injected at such a high rate of speed it is important that the proper size of IV catheter is used and that the catheter is placed in a vessel that can accommodate the pressure of the injection.
For CTA exams it is important that the IV catheter not be any smaller than a 20g. If the selected vein can handle an 18g catheter, that would be preferred if performing a CTA exam.
Due to the high injection rate and the small vessels in the hand, it is not recommended to use a hand IV for CTA exams.
For most contrast exams and especially CTA exams, it is ideal to use the right side for the placement of the IV if possible.
Look at your anatomy book and compare how the right and left side connect to the superior vena cava, to better understand why the left side would put off more artifact than the right when performing upper body CTA exams.
Scan Timing – CT Training – CTA Exam
The critical part of a CTA exam is to correctly time the start of the scan to occur when the contrast density is at its peak in the anatomy of interest.
Scan timing can be accomplished in two different ways:
1) One way is with the use of an ROI (Region of Interest) and special software that is designed to monitor the contrast density within a vessel and trigger the scan automatically when the set contrast density threshold is met. (For more information, see ROI in CT Basics).
The name for the ROI contrast monitoring function varies by manufacturer.
GE: Smart Prep, Toshiba: Sure Start, Siemens: Care Bolus, Philips: Bolus Pro.
When the exam is started and the scouts are set up and scanned, a special axial scout image is done in the area that the ROI will be placed. In Figure 4-5 the light lines represent the AP scout of the entire anatomy of interest and the dark line at the tip of the arrow represents the extra axial scan to look at the anatomy for the placement of the ROI.

Looking at Figure 4-6 we see the extra axial scout image with the ROI placed within the pulmonary artery. As the scan is being set up, the technologist will enter a Hounsfield number into the scanner software. This number tells the system that when the Hounsfield number of the ROI reaches the threshold set by the technologist, to automatically start the scan.

When the scan parameters are set and the ROI is placed, confirm your settings, and press the contrast injection button and the scan button at the same time. The scanner will now continue to scan in the ROI, measuring the density within the vessel. Once the contrast reaches the ROI and the density of the contrast is high enough to trigger the set threshold, the scan will start.
As technology has improved, the use of an ROI tends to be the most common way to properly time CTA exams.
NOTE: In the Virtual CT Trainer software the previously explained CTA timing concept is used for the PE study and Neck CTA (Carotids) but can be used for any CTA exam on a real scanner. Practice this concept as much as you need too and get comfortable with how this works in the CT environment. (For more details see the PE Study and Neck CTA exam in the Step by Step CT scanning procedures).
2) Another way to determine the timing for a CTA study is by performing a test injection and then scanning the patient in the same location multiple times one second apart.
After all the scan parameters are set and the scouts have been acquired, you will now position a line on the AP scout in an area that is in the proximity of the desired CTA anatomy (ours is the line at the bottom of the mandible in Figure 4-7). The protocol will then be set to scan in the exact location 24 times with a one second rotation time (see your department protocol).

Set up your injector to inject about 10-15 cc of contrast using the injection rate set for your main CTA exam. Start the 24 scans and the injector at the same time. When the scans have completed you will have 24 images spaced one second apart. These 24 images will be used to help determine how many seconds it took for the contrast to reach the area of interest.
Looking at the 24 images in Figure 4-8 you begin to see the contrast in the carotid arteries on image 16, but the contrast does not become bright until image 18. Seeing the contrast bright on image 18 tells you that it took 18 seconds for the contrast density to be sufficient for a quality CTA exam.
When setting up your main CTA scan, enter 18 as the scan delay and confirm your settings. Some technologists will add 1 – 2 seconds to this number to ensure a bright contrast enhancement. Now start the scan and the contrast injection at the same time, scanning through the anatomy of interest.
NOTE: In the Virtual CT Trainer software, the CTA timing concept is used for the Head CTA (Circle of Willis (C.O.W)), practice as much as you need to and get comfortable with how this works in the CT environment. (See the head CTA exam in the Step-by-Step CT scanning procedures for additional information).
Post Processing

Most CT departments will require the technologist to do some post processing (3D, MIPS) on CTA exams.
Depending on your department protocol, the additional images can be processed using the built-in scanner software, but many times the images are sent to a separate workstation with special software for post processing (see Figures 4-1, 4-2, 4-3 and 4-4 for post processed images).
End of CTA Exams
CT Training Navigation
Prev “CT Number” Next Coming “Artifacts”
How to Get Cross-Trained in CT – Computed Tomography Training
Most Popular Courses



All Our X Ray CE Credits Satisfy ARRT® CE Requirements – ARRT®
Disclaimer: This training does not replace formal hands-on CT training, either on the job or through academic schooling. Do not perform CT on an actual patient without additional training.
Activities on this site may be available in multiple formats or from different sponsors. A CE activity can be completed only once per biennium.
ARRT®, American Registry of Radiologic Technologist®, CQR®, Structured Education®, R.R.A.®, R.T.®, ASRT®, American Society of Radiologic Technologist®, ARDMS®, NMTCB®, and AHRA® are registered trademarks owned by their respective companies. Rad Tech Education LLC is independently owned and is not licensed or endorsed by these companies.