CT Training – CT Basics 2
CT Training – CT BASICS 2

Scanning the patient
After laying the patient on the table and positioning them by using the internal or external laser lights (see CT basics 1 Figure 1-4), you will zero out your scanner. The zero location is the point the scanner will begin the first scout image.
Your department CT scanner will have built-in protocols that will match the type of exam you are performing (see sample written protocol – Section 21). The protocols should be set to acquire the images required by the radiologist and optimized to minimize patient radiation dose, while maintaining image quality.
The scanner techniques will be generally set for the standard patient and will need to be adjusted to accommodate a different body habitus. Most facilities will have separate protocols for pediatric patients that are set up specifically to reduce radiation dose as well as increase the speed of the scan to reduce motion artifact (see the pediatric patient – Section 8).
Scout (Topogram) Images – CT Training – CT BASICS 2
Figure 2-1


NOTE: If you place the patient on the table in a different position than is set in the protocol, such as: head first, feet first, supine or prone, you will need to adjust the protocol to compensate for the new patient position. After changing the patient’s position in the protocol, the computers are now able to mark the proper right or left side on the images. It is also important to look at the direction of the table movement set in the protocol.
Double check that the setting in the protocol matches the direction you want the patient to move (in or out of the gantry).
When performing the AP scout image the x-ray tube will be above the patient with the receptors directly below. The patient will move through the x-ray beam and acquire an AP anatomical image similar to a standard x-ray (see Figure 2-1).
Figure 2-2


When performing the lateral scout image, the x-ray tube and receptors will rotate to a lateral cross table position, to acquire the lateral scout image as seen in (Figure 2-2).
Depending on the manufacturer, some scanners will obtain the AP scout and require the patient to move back to the original starting point of zero (0) to begin the lateral scout scan. Other scanners will move the patient to acquire the AP scout and then acquire the lateral scout as the patient moves back into the scanner to the starting point of zero (0).
Conventional CT (Step-and-Shoot) was the typical mode of scanning prior to the introduction of the slip ring. In the early days of CT, scanners used wires/cables to provide the power to the x-ray tube and transmit the data from the detectors. The problem with this design was, the tube/detector movement was limited to one 360° rotation before it had to return back to the start position. This proved to be a problem with chest and abdominal scans due to the length of the scan and the requirement of multiple breath holds to get through the entire anatomy.
During the scanning process, the tube and detector would rotate 360° around the patient while the table remained stationary. The table would then move to the next slice location as the tube and detector rotated back to their start position. The scanner would then continue this process until all of the set slice locations were completed.
Helical CT – CT Training – CT BASICS 2
Helical CT has replaced conventional (Step-and-Shoot) CT in most cases, but conventional CT may still be used in protocols that require a higher level of detail, such as the internal auditory canals (IAC).
Helical CT (see Figure 2-3) is used in the majority of modern day protocols and has allowed for imaging abilities that were not available with conventional CT.
Figure 2-3

The invention of the slip ring and brushes has allowed for a continuous 360° rotation of the tube and detectors without the need for wires/cables to be connected.
As the tube rotates, the table will continuously move the patient through the gantry/x-ray beam allowing for the continuous acquisition of raw data (see raw data under image reconstruction – Section 18). The speed of the table as it moves the patient through the gantry is determined by the pitch setting in the protocol (see pitch below).
MA (millamperage) or tube current represents the quantity of photons emitted from the x-ray tube. When the MA setting in the software is combined with the rotation time (time it takes for the x-ray tube to make a full rotation around the gantry) you get MAs (MA x Time = MAs). CT uses the same concept of MAs that is used in general x-ray just in a higher volume.
Image Noise (Grainy Image) – CT Training – CT BASICS 2
NOTE: A CT that is scanned with the MAs set too low for the anatomical density, will have a grainy appearance (noise) due to a decreased number of x-ray photons reaching the detectors (see Figure 2-4).
Many modern CT scanners have a feature called MA modulation. The purpose of MA modulation is to adjust the MA settings (which will change the overall MAs) as it is scanning through the patient. Rather than using the same MA settings throughout the entire exam it will adjust the MA up or down as needed to help reduce the radiation dose to the patient.
Figure 2-4

The MA modulation software determines areas of the body that have more or less density and adjusts the MA (MAs) accordingly (for more information on MA modulation, see radiation safety.
KV (Kilovoltage) or tube voltage represents the energy or penetrating power of the X-ray photons. KV is used the same way in CT as it is used in general radiology and determines the energy of the x-ray photon and its ability to get through dense areas of the body.
NOTE: Most CT scans will use a standard KV setting of 120 to 140 kv and is not typically a parameter that used to adjust for a different body habitus.
MAs is the combination of the MA setting and the exposure time or x-ray tube rotation time to determine a MAs value (MA x Time = MAs).
Pitch = amount of table movement per 360° rotation of the tube, divided by the slice thickness or the thickness of collimated beam.
CT Table Pitch – CT Training – CT BASICS 2
As the pitch is increased the table will move faster through the gantry/x-ray beam, but as a tradeoff results in a decrease in image quality. Looking at the slinky in Figure 2-5 and Figure 2-6 (helical CT) we see that the faster the patient moves through the gantry the more stretched the slinky (helical) becomes (Figure 2-6).
Sometimes in the case of an uncooperative child or a trauma patient, scan speed is of the essence and the decrease in image quality no longer becomes an issue.
If the scan speed and image quality are both an important factor, you can compensate for the loss in image quality by increasing the MAs, but as we know, an increase in the MAs results in a higher radiation dose to the patient.
Figure 2-5

Figure 2-6

Increasing the pitch can be especially useful with the pediatric patient. An increase in pitch will allow you to cover more anatomy in a shorter period of time and reduce the chance of motion. If all other settings remain the same, increasing the pitch reduces the radiation dose, which becomes increasingly important with the pediatric patient.
An increase in pitch also becomes helpful in CTA exams when scan speeds become an important factor for a quality CTA exam (see CTA imaging). CT Training – CT BASICS 2
ROI (Region of Interest) is a measuring device that measures the Hounsfield number of a specific anatomical area the ROI has been placed over. When an ROI is placed over a suspicious area in the anatomy, a Hounsfield number (HU) is displayed (see Figure 2-7) that will indicate to the radiologist if it is cystic in nature or a solid mass. For more information about Hounsfield numbers (HU) see CT numbers.
ROI Measurements – CT Training – CT BASICS 2
The ROI can also be placed within a vessel to measure the contrast density for a CTA study. To use an ROI for a CTA study, the technologist will place the ROI in an artery near the anatomy of interest and set a threshold (trigger point) for the ROI. When the contrast density within the vessel reaches the threshold set by the technologist, the scan will start automatically. This technique allows for the scan to be started while the contrast enhancement within the vessels of interest is at their peak. See the CTA imaging
Figure 2-7

PACS (picture archiving and communication system) is used to store and retrieve digital medical images for the use of diagnosis and treatment of medical conditions. CT Training – CT BASICS 2
Measuring features are built into the CT scanner software and are also available on your PACS system. Measurements of suspicious areas are used to determine the size of an abnormality seen in the anatomy and can also aid in the diagnosis and treatment of the patient (see Figure 2-8).
Figure 2-8

Timing of the contrast injection is very important in aiding in the ability to see an abnormality. Some pathology may not be visualized by the radiologist if the contrast material was not administered or the patient was scanned at an improper time (see common delay times below). The injection of contrast material at specific times (multi-phase study) can also aid the radiologist in determining what type of pathology is present and if it is benign or malignant.
Common Scan Delay Times
Common delay times (subject to your department protocols)
- Abdomen/Pelvis – 70 second delay
- Chest – 25 to 30 second delay
- Soft Tissue Neck – 40 second delay
- Standard brain – 3 to 5 minutes
- Multiphase exam
- Arterial phase 25 to 30 seconds
- Venous phase 60 to 70 seconds
- Delay phase 3 to 5 minutes
- CTA exams – Timed using ROI or timing bolus test (See CTA Imaging)
End of CT Basics 2
CT Training Navigation
Prev “CT Basics 1” – Next is “CT Number“
How to Get Cross-Trained in CT – Computed Tomography Training
Most Popular Courses



All Our X Ray CE Credits Satisfy ARRT® CE Requirements – ARRT®
Disclaimer: This training does not replace formal hands on CT training, either on the job or academic schooling. Do not perform CT on an actual patient without additional training.
Activities on this site may be available in multiple formats or from different sponsors. A CE activity can be completed only once per biennium.
ARRT®, American Registry of Radiologic Technologist®, CQR®, Structured Education®, R.R.A.®, R.T.® ASRT®, American Society of Radiologic Technologist®, ARDMS®, NMTCB®, AHRA®, Are Registered Trademarks Owned by Their Respected Companies. Rad Tech Education LLC is Independently Owned and is Not Licensed or Endorsed by These Companies.