CT Training – CT Basics 1
CT BASICS 1

CT has made some dramatic advancement over the last several years and has become an excellent diagnostic tool in modern day health care.
In this section we are going to look at some of the basics of CT to help familiarize you with the CT equipment and setting up the scanner for a patient.
First let’s take a look at the some of the CT hardware that is important to you as a technologist.
Looking at Figure 1-1, these are some of the basic scanner components that you will need to be familiar with as a CT Technologist.
The Gantry – CT Training – CT Basics 1
Figure 1-1

- Gantry – The gantry is what houses many of the parts used to make a CT scan possible. In the gantry you will find the x-ray tube, filters, collimators and image receptors, among many other parts that help to produce a CT image. When scanning, the parts inside the gantry rotate (360°) around the patient at a high rate of speed collecting raw data as the table/couch moves the patient slowly through the gantry/x-ray beam.
- Table/Couch – The CT table is referred to as the “couch” by most manufacturers. However in the radiology field you will hear the term table used most often, although either can be used interchangeably. The table will come with two different attachments that can be easily changed for different scans (see Figure 1-1 and Figure 1-2). The head holder will typically be used for head and neck scans (with the exception of trauma) and the footboard will be used for all other types of exams. During a CT scan the patient will lay on the table as they slowly move through the gantry/x-ray beam with the x-ray tube and receptors rotating around them. The table movement will begin at the point where you set your first slice using your scout images, and will continue moving until the table reaches the table location of the last slice set on your scout images. When doing a helical scan, the speed of the table as it moves the patient through the gantry/x-ray beam, is determined by the pitch (see helical CT In CT basics 2). The pitch is typically preset in the scan protocol and in most cases does not need to be adjusted. If the technologist is faced with a difficult patient or an emergency situation he/she may choose to increase the pitch to achieve a faster scan time if needed.
Figure 1-2

- Gantry controls – (see below and Figure 1-3 for more details)
- Laser lights – The horizontal/side laser lights are used to position the anatomy in the center of the scan bore. The internal/external centering lights (see internal/external zero below) are used to set the starting anatomical point of the scout scan.
- Foot Petals – Most scanners will be equipped with foot petals that control the table movement up and down and also move the table in and out of the gantry. The foot petals are there as a convenience for the technologist and are not required to be used while positioning the patient. All of the same functions that can be done using the foot pedals can also be done using the gantry control panel.
- Injector – The injector is what is used to administer the contrast material to the patient through an IV line at a set rate of speed. Once the injection parameters have been set and the injector has been armed, the injector can be started from within the scan room or by using the injector controls in the scanner control room. See pre scanning procedure – Section7 for more information on drawing up contrast and setting up the injector.
Figure 1-3

In Figure 1-3 we will look at some common gantry controls that are needed to position the patient correctly. Depending on your scanner manufacturer, the layout may be different but should include the basic controls.
Get familiar with your facilities scanner controls.
If you are not able to obtain access to your CT suite, find out the brand name and model of your department scanner and look up the control layout on the internet. This will allow you to be comfortable with the scanner controls when you do get the opportunity to access the CT suite.
- Table height – When the patient is secure and comfortable on the table you can raise the table height using the table height buttons or foot pedals. Adjust the table height so that the horizontal laser light is midway between the anatomy of interest.
- Laser light on/off – Some scanners will turn the laser light on for you as you move the patient in to the gantry. If not, simply turn on the laser light to position the patient for the scan you are performing. Avoid allowing the patient to look directly into the laser lights as you position them. If your scanner does not turn off the laser light when the scanner is placed to zero, be sure to manually turn off the laser light.
- Table in/out – These buttons allow you to move the table in and out of the gantry as needed for positioning or to bring the patient out of the gantry after the scan is completed. Most scanners will have a slow and fast speed for moving the table in and out of the gantry.
- Internal/External zero – When you are centering the patient for the scan, you will see two centering lights that go across the anatomy of interest (see Figure 1-4). One is the external light and the other is the internal light. Place one of these lights just above the area of interest and zero out the scanner. Where you zero out, tells the scanner where on the patients anatomy you want to start your scout image.
Figure 1-4

- Gantry tilt/angle – After you have done your scout images you will place your lines to cover the desired anatomy. If the scan you are performing requires an angle of the gantry, you will place that angle using the lines on the scout. After you have confirmed your line placement, if there was an angle placed at the time the lines were set up, the scanner will require you to angle the gantry to match the angle that was placed on the scout images. Most new scanners will have a gantry angle button on the scanner keyboard (see Figure 1-5) that allows you to angle the gantry from the control room.
If your scanner does not have angling capabilities from the control room, you will need to manually angle the gantry using the gantry control panel.
Figure 1-5 is an image of a scanner keyboard. The keyboard layout will vary with different manufacturers and the age of your scanner.
If you are not able to obtain access to your CT suite, find out the brand name and model of your department scanner and look up the keyboard layout on the internet. Try to get familiar with the scanner keyboard in your facility.
Figure 1-5

- Scan/prep button – After you have set up your scout images and have pressed confirm, your scan/prep button will start flashing. Press the button once to prep the scanner. After the scan/prep button is pressed once it will turn to a solid color and will no longer be flashing. Once the scan/prep button is a solid color, the scanner is prepped and ready to start the scan. The scan/prep button should be similar on most manufacturers but will vary.
- Talk button – This button is used to verbally communicate with the patient without having to go out to the scan room. The gantry in the scan room and the keyboard in the control room will have a speaker and microphone, allowing for two way communication between the technologist and patient at all times during the scan. Press the talk button to speak to the patient and release the talk button to hear the patient.
- Scan abort button – The scan abort button will immediately stop the scanning process. This is typically done if the patient is having a contrast reaction, extravasation of the contrast material or the patient is not holding still on the scan table.
- Emergency stop button – The emergency stop button will instantly shut off power to the scanner in case of an emergency situation. There should be an emergency stop button located on the keyboard as well as on the gantry out in the scan room.
Scan Setup
Figure 1-6 was taken from the head CTA exam within the Virtual CT Scanner software. Using the head CTA exam, we will teach you how to manually determine the delay time that is needed for a quality CTA exam.
Figure 1-6

Beginning represents the table location in millimeters (mm) where the scan will start.
End represents the table location in millimeters (mm) where the scan will end.
Image Count is the number of images that will be acquired during the scan and is determined by the distance of the scan from the beginning to the end and how thick each slice will be. Line number 2 in Figure 1-6 starts at 19 mm from where the scanner was zeroed out and ends the scan at 142 mm. This is a scan distance of 123 mm with a slice thickness of .5 mm resulting in an image count of 246.
Slice Thickness represents the acquired image thickness at the time the patient is scanned and does not represent the thickness of the final images after they have been reformatted. Typically images will be acquired using a thin slice thickness of .5 – 1 mm and will be reconstructed at a thicker slice thickness such as 3 – 5 mm.
Gantry angle represents the amount of angle that is placed on the gantry.
Scan Time is the amount of time it takes to scan the selected anatomy. The scan time is controlled by many different variables, such as: amount of anatomy covered, acquired slice thickness, rotation time and the amount of pitch used.
Rotation Time is the amount of time in seconds it takes for the x-ray tube to make one full 360° rotation around the gantry. Rotation Time + MA Setting = MAs
Scan Timing
Looking at line 1 in Figure 1-6 we can see that the beginning and ending location are both fifteen (15), with a total of 24 images acquired using a one (1) second rotation time. What this scan setup accomplishes is that we now have 24 images in the same anatomical location that are spaced one (1) second apart.
The purpose of this scan setup is to inject a small amount of contrast at the same time the scan of 24 images is started. After the scanning is complete we can look at the 24 images and determine what image begins to show bright contrast in the artery of interest. We know from our statement earlier that each image is spaced one (1) second apart and represents one (1) second of time. If we determine image 18 (as in the Head CTA exam) has bright contrast enhancement, we know that it took eighteen (18) seconds for a high density of contrast to reach that anatomical location. This information allows us to set the proper delay time to start the scan, when the contrast enhancement is at its peak.
NOTE: Many technologists when using this technique as a guide for contrast timing will add one (1) second or two (2) seconds to ensure good contrast enhancement when the scan is performed.
For more information on scan timing, see CTA exams
End of CT Basics 1
CT Training Navigation
CT Basics 1 – Next “CT Basics 2“
How to Get Cross-Trained in CT – Computed Tomography Training
Most Popular Courses



All Our X Ray CE Credits Satisfy ARRT® CE Requirements – ARRT®
Disclaimer: This training does not replace formal hands on CT training, either on the job or academic schooling. Do not perform CT on an actual patient without additional training.
Activities on this site may be available in multiple formats or from different sponsors. A CE activity can be completed only once per biennium.
ARRT®, American Registry of Radiologic Technologist®, CQR®, Structured Education®, R.R.A.®, R.T.® ASRT®, American Society of Radiologic Technologist®, ARDMS®, NMTCB®, AHRA®, Are Registered Trademarks Owned by Their Respected Companies. Rad Tech Education LLC is Independently Owned and is Not Licensed or Endorsed by These Companies.