CT Artifacts – Learn CT at Home
CT Artifacts
Common CT Artifacts
Motion Artifact – CT Artifacts
If a patient is poorly educated on what will happen during the scan or how they may feel when the contrast is injected, you drastically increase your chances for voluntary patient motion. When the patient has
been properly informed about breath holds, table movements and contrast sensations, they will be prepared for them which in turn decrease the chances for motion on the final images.
Sometimes motion is unavoidable such as: bowels moving (peristalsis),
patient has an uncontrollable tremor, scared child (see pediatric patient); patient may be unable to hold their breath the entire scan or the inability to communicate properly due to a language barrier.
Do what you can to reduce motion by properly informing the patient about the scan, use the straps/restraints that were supplied with the scanner and use an interpreter if available for a language barrier.

Beam Hardening Artifact
Beam hardening artifact happens as the x-ray photons pass through thicker bony parts of the body like the shoulders/upper chest, hips or back of the skull. As the photons pass though these areas, the body absorbs the lower energy photons, leaving only the higher energy photons to continue through the anatomy and reach the detectors.
When scanning through dense anatomy or contrast material, you will see lighter or darker streaking across the area where the beam was hardened from lower energy photons being absorbed.

Out of Field Artifact
Out of field artifact or (truncation artifact) is caused when part of the patient’s anatomy is outside of the scan field of view. The out of field artifact will appear as streaking or shading, as seen in the outer edges of the image. Along with streaking or shading, it can produce a bright band or halo near the image edges, obscuring peripheral anatomy and causing inaccurate CT numbers.
The artifact happens when anatomy extends beyond the scan field of view, leaving incomplete projection data for reconstruction. It commonly occurs with larger patients or patients positioned off-center.
To avoid Center the patient carefully, reposition the arms when appropriate, and select a scan field of view large enough to include the patient’s anatomy

Ring Artifact
The ring artifact is caused by a detector element that has failed or is out of calibration.
One or more bright or dark rings centered on the scanner’s axis of rotation, most easily recognized on axial images.
An affected detector element consistently records an incorrect signal as the tube and detectors rotate. Reconstruction turns that repeated error into a circular pattern.

Partial Volume Artifact
The partial volume artifact is an inherent part of the CT modality, but we can reduce it by using thinner slices and smaller voxels.
Looking at the image, a voxel contains a chunk of anatomical data that is equal to the thickness of that slice. Think of a voxel that starts at the
base of the lungs and as the slice thickness is increased the voxel depth
goes deeper into the lower tissues, such as the diaphragm or upper abdominal cavity. Now this new long voxel contains data from the lung tissue where it started and also contains data from the abdominal tissue where it ends, as well as any tissues in between. When the software reconstructs the images with the new slice thickness, it will average both the lung tissue data and the abdominal tissue data that are
now in the voxel together. When the averaging is complete it will
assign the new pixel shade of grey that is equal to a weighted average of both tissue types (see CT numbers). This new pixel that was created is no longer a lung pixel or an abdominal pixel but a combination of the two and does not give a true representation of either tissue type.
This effect of partial volume averaging can cause the visible image to
appear fuzzy (blurred tissue boundaries and reduced visibility of small structures). Partial volume averaging can also have the appearance of false pathology or it can hide existing pathology, due to the averaging of different tissue types.
Our biggest defense against partial volume artifact is to keep the slice
thickness as small as possible, but thinner slices also causes image noise (making the images look grainier) if no other parameters are adjusted. So there needs to be a balance of decreasing the overall slice thickness and image quality without increasing the dose to the patient.

Metal Artifact
Metal artifact is a common CT artifact that can appear as bright and dark streaks extending from a metal object, sometimes obscuring nearby anatomy (see Figure image). Sources include jewelry, metal on the patient’s clothing, objects in their pockets, and internal metal such as dental restorations, pacemakers, surgical clips, and orthopedic implants.
Metal strongly attenuates the x-ray beam, causing beam hardening and photon starvation, which contribute to the streaking. Removing external metal from the scan area before imaging helps prevent this artifact. When internal metal is present, dedicated metal artifact reduction software can reduce the streaking, although it may not eliminate it completely.

CE Credits



All Our X Ray CE Credits Satisfy ARRT® CE Requirements – ARRT®
Activities on this site may be available in multiple formats or from different sponsors. A CE activity can be completed only once per biennium.
ARRT®, American Registry of Radiologic Technologist®, CQR®, Structured Education®, R.R.A.®, R.T.® ASRT®, American Society of Radiologic Technologist®, ARDMS®, NMTCB®, AHRA®, Are Registered Trademarks Owned by Their Respected Companies. Rad Tech Education LLC is Independently Owned and is Not Licensed or Endorsed by These Companies.